MAC CAT-1006 FINAL

Roundtest

Roundtest RA-120/120P SERIES 211 — Roundness Measuring Instruments

Easy operation, compact and outstanding cost/performance ratio, designed for use on the shop-floor right beside the production line. • D.A.T. (Digimatic Adjustment Table) function aids adjustments such as centering and leveling, and substantially reduces the time required for preliminary setup operations.

What is the D.A.T. function? <Patented>

Initial misalignment of axes

Turntable axis

Turntable axis

Workpiece

Workpiece

Initial misalignment of centers at turntable surface

Feed the digital micrometer head of the trntable by the deviation displayed.

Centering range: ±3mm Leveling range: ±1°

Simple adjustment

Centering / leveling complete

Preliminary measurement

Display of misalignments

Preliminary measurement of two cross-sections ’A’ and ‘B’.

Dedicated analysis unit type (RA-120)

Data analysis by PC (RA-120P)

RA-120

•  ROUNDPAK , a data analysis program employs Windows OS and archives higher level of analysis.

• Compact, lightweight design due to incorporating electronic components inside the main unit.

Compact Roundness Measurement ROUNDTEST RA-120/120P

The analysis capabilities for the various models ( RA-120 / 120P / 10 ) vary. For details, refer to page L-26. RA-120P

SPECIFICATIONS Model No.

J

Catalog No.E15008(4)

RA-120

RA-120P

Refer to the Roundtest RA-120/120P Brochure ( E15008 ) for more details.

211-544-13 211-543-13 211-547-11 211-546-13

inch/mm

-

-

DAT Function Radial direction

Installed

Installed

(0.04 + 6H/10000) µm H: Probing height (mm) Axial direction (0.04 + 6X/10000) µm X: distance from the center of rotation (mm)

Rotational accuracy (JIS B 7451-1997)

ø280 mm (ø380 mm: for the vertical position when detector holder is installed reversely, the maximum probing height is up to 50 mm from the table top.)

Turntable

Maximum probing diameter* 1

Maximum loading mass

25 kg

Vertical movement

Vertical travel

280 mm

X-axis

Travel range

165 mm (-25 mm to 140 mm from the rotation center)

Detector* 2 ±1000 µm *1 Auxiliary stage for a low-height workpiece (optional) is required for the measurement 20 mm or less in the radial direction from the center point of the table and 20 mm or less from the table top. *2 Only the standard length stylus is applicable to this detector. The long type cannot be used. Measuring range

Get a Quote

J-21

Powered by