Laser Scan Micrometer
Optional Accessories • Multifunctional display unit, LSM-6200 :
LSM-500S Measuring Unit SERIES 544 — 5 μm to 2 mm Measuring Unit
Order No. Remarks 542-072A English mm/in English user's manual Display type
• Capable of measuring down to 5 µm outside diameter. • Provides ultra-high accuracy of ±0.3 µm over the entire measuring range (5 µm to 2 mm).
• Panel-mount type display unit, LSM-5200 :
Order No. 544-047
Remarks
English user's manual
SPECIFICATIONS Order No. (Laser only)
• Standard calibration gage set (ø0.1, ø2.0): 02AGD110 • Guide pulley: 02AGD200 • Air blower: 02AGD220 • Extension signal cable (max. 15 m)
544-532
Package No. (with LSM 6200 Display)
64PKA117
Applicable laser standards
IEC, FDA
User's manual Measuring range
English version
Order No. 02AGN780A 02AGN780B 02AGN780C
Cable length
.0002" to .080" (0.005 to 2mm)* 1
5 m 10 m 15 m
Resolution
.000001" to .0005" (0.01 to 10μm) (selectable)
Repeatability* 2 Accuracy (20°C)* 3 Positional error* 4 Measuring area* 5 Laser wavelength Laser scanning speed Scanning rate
±0.03μm ±0.3μm ±0.4μm
1×2mm (0.005 to 2mm)
3200 scans/s 650nm (Visible)
76m/s
Temperature
0 to 40°C
Operating environment
Humidity
RH 35 to 85% (no condensation)
Protection Level IP64* 6 *1: The measuring range for the transparent object will be 0.05mm to 2mm. Please consult your local Mitutoyo office for objects smaller than 0.05mm. The measuring range will be 0.1mm to 2mm in the 1 to 255 edge measurement mode or when activating the automatic workpiece detection. If using the optional dual-connection unit for LSM-6200, the measuring range will be 0.05mm to 2mm. *2: Determined by the value of ±2 ( : standard deviation) when measuring ø2mm at the interval of 0.32 sec. (average 1024 times). *3: Center of the measuring range for cylindrical workpieces outside diameter. *4: An error of the outside diameter due to variation in workpiece position either in the optical axis direction or in the scanning direction. *5: The area given by [optical axis direction]×[scanning direction]. *6: If the workpiece or glass of the measuring unit window is soiled by water or dust, the unit may malfunction. Note: When using extra-fine line measurement function (FINE), guide messages for setting the following will not be displayed: dual-measurement, segment designation, automatic workpiece detection and group judgment.
Schedule Service
G
LSM-501S Measuring Unit SERIES 544 — 50 μm to 10 mm Measuring Unit
• Provides ultra-high accuracy of ±0.5 µm over the entire measuring range (0.05 to 10 mm). • The industry’s first narrow-range accuracy performance in this measuring range of ±(0.3+0.1 D) μm is available for high-accuracy measurement.
Optional Accessories • Multifunctional display unit, LSM-6200 :
Order No. Remarks 542-072A English mm/in English user's manual Display type
• Panel-mount type display unit, LSM-5200 :
SPECIFICATIONS Order No. (Laser only)
Order No. 544-047
Remarks
English user's manual
544-534
Package No. (Laser w/LSM 6200 display)
64PKA118
Applicable laser standards
IEC, FDA
• Standard calibration gage set (ø0.1, ø10.0): 02AGD120 • Wire guiding pulley: 02AGD210 • Adjustable workstage: 02AGD400
User's manual Measuring range
English version
.002" to .4" (0.05 to 10mm)
Resolution
.000001" to .0005" (0.01 to 10μm) (selectable)
• Air blower: 02AGD230 • Workstage: 0 2AGD270 • Extension signal cable (max. 15 m)
Repeatability* 1
±0.04μm ±0.5μm
Whole range Small range
Accuracy* 2 (20°C) Positional error* 4 Measuring area* 5 Scanning rate Laser wavelength Laser scanning speed
±(0.3+0.1 D)μm* 3
±0.5μm
Order No.
Cable length
2×10mm (ø0.05 to ø0.1mm) 4×10mm (ø0.1 to ø10mm)
02AGN780A 02AGN780B 02AGN780C
5 m 10 m 15 m
3200 scans/s 650nm (Visible)
113m/s 0 to 40°C
Temperature
Operating environment Protection Level
Humidity
RH 35 to 85% (no condensation)
• Extension relay cable Order No.
IP64* 6 *1: Determined by the value of ±2 ( : standard deviation) when measuring ø10mm at the interval of 0.32 sec. (average 1024 times). *2: Center of the measuring range for cylindrical workpieces outside diameter. *3: D=Difference in diameter between the master gage and workpiece (Unit: mm) *4: An error of the outside diameter due to variation in workpiece position either in the optical axis direction or in the scanning direction. *5: The area given by [optical axis direction]×[scanning direction]. *6: The protection level provided for the interior. If the workpiece or glass of the measuring unit window is soiled by water or dust, the unit may malfunction.
Cable length