MAC CAT-1006 FINAL

Mu-checker

Lever/Cartridge Probe Heads SERIES 519 — Electronic Micrometer

SPECIFICATIONS Lever heads Model

Common specifications • Connection: Half-bridge • Cable length: 2 m • Connector type: MAS-5100 (DIN5P) or equivalent

MLH-521 519-521

MLH-522 519-522

MLH-326 519-326

MLH-327 519-327

Order No.

Measuring range (mm)

±0.5

Stroke (mm)

±0.6

±0.65

Measuring force (N)

Approx. 0.2

Approx. 0.02

Approx. 0.15

Linearity (%) Stylus support

±0.3

±0.5

Pivot bearing

Pivot bearing

Parallel-leaf spring

Pivot bearing

Note: A ø2 mm ball-ended stylus is supplied as standard with all probes.

519-521 MLH-521

519-522 MLH-522

• Interchangeable styli: ø1: 520940

• Interchangeable styli: ø1: 520940

(Standard accessory)

(Standard accessory)

ø2: 520939

ø2: 520939

(Standard equipment)

(Standard equipment)

ø3: 520938

ø3: 520938

(Standard accessory)

(Standard accessory)

ø 6.5

21.3

ø 14

9.2

35 19

ø 6.5 ø 14

21.3

19

9.2

35

ø 6.6

ø 6.6

24.9

5.8

24.9

5.8

13

30.7

52.3

30.7

52.3

13

Unit: mm

Unit: mm

G

519-326 MLH-326

519-327 MLH-327

• Interchangeable styli: ø1: 102824 (Optional) ø2: 102825

• Interchangeable styli: ø1: 102824 (Optional) ø2: 102825

(Standard equipment)

(Standard equipment)

L=34 エクステンションバー装着時(同梱品)

ø3: 102826 (Optional) • Extension bar

ø3: 102826

6.5 34 7

(Optional)

25

24

ø6.5 ø14

L=9 mm (Standard equipment)

ø6.5 ø14

(25)

9 7

32.3

6.5

6.5 15.8

L=34 mm (When a bar for extending the distance to the stylus tip is supplied)

25.3

56

13

22.5

5.5

24.6

43

13

Unit: mm

Unit: mm

Transfer Stand

Note on stylus angle If the stylus of a pivot bearing type probe makes an angle with a workpiece surface, as in the figure, calibration should be performed for accurate measurement. Alternatively, the displayed value may be corrected by multiplying it by the appropriate correction factor as given in the table. Model 519-326 does not need correction.

Dimensions of dovetail plate on probe body Enables mounting on a lever head mounting bracket or stem.

Unit: mm

ø30

80

Stem mounting hole without Bush: ø9.53 with Bush: ø8

Platen Parallelism: 1.3 µm or less

Unit: mm

ø58

60 ゜ 15

Angle ( θ ) Correction factor 0° 1.00 10° 0.98 20° 0.94 30° 0.87 40° 0.77 50° 0.64 60° 0.50

( 5.934 )

519-109-10 (with a serrated plate)

168

Main Specifications

θ

Effective transfer range (mm)

Fine adjustment range (mm)

Mounting hole (mm)

Order No.

Without Bush: ø9.53 With Bush: ø8

519-109-10

0 - 320

1

Display value × Correction factor = Corrected value

G-22

Powered by