MAC CAT-1006 FINAL

Litematic

VL-50-B/50S-B Litematic SERIES 318 — High-accuracy/resolution Measuring Machine

Optional Stand for VL-50S-B

• With a measuring force of only 0.01 N, the Litematic is ideal for measuring easily deformed workpieces or high-accuracy components. • For workpieces for which 0.01 N is insufficient, either the 0.15 N or 1 N model is recommended. • The motor-driven spindle moves up/down and stops when the contact point touches the workpiece. Then the maximum, minimum and runout values are measured under a constant force.

• High resolution of 0.01 µm, and wide measuring range of 50 mm.

Unit: mm

957460

• Measuring system VL-50-B , integrated display type, and VL-50S-B , a separate display type, are available. • The measuring table supplied with VL-50-B is ceramic, which is corrosion free, for easier maintenance and storage. • The spindle is made of low thermal expansion material.

ø50

Litematic head

Column

Height adjustment nut Base Bracket Hex socket set screw (4-M4×0.7)

157

48

80

318-221-10A

318-226-10A

Clamp screw

300

SPECIFICATIONS Model

Shown with optional granite base, 957460

VL-50-B VL-50-15-B VL-50-100-B VL-50S-B VL-50S-15-B VL-50S-100-B 318-221-10A 318-222-10A 318-223-10A 318-226-10A 318-227-10A 318-228-10A

Order No.

Optional Accessories • Foot switch: 937179T

Measuring range

0 to 50 mm (0 to 2 in)

G

Resolution Display unit Scale type

0.01/0.1/1.0 µm (0.0000005 in/0.000005 in/0.00005 in) 8 digits/14 mm (0.6 in) character height (without signs)

• Dedicated stand: 957460 * 1 • SPC cable (1 m): 936937 * 2 • SPC cable (2 m): 965014 * 2 • VL weight part: 02AZE375 * 3 • Recommended contact points: Shell type: 101118 (Approx. 0.02 N)* 4 Carbide tipped spherical contact point, ø7.5: 120059 (Approx. 0.03 N)* 4 Carbide tipped spherical contact point, ø10.5: 120060 (Approx. 0.06 N)* 4 Carbide tipped needle contact point, ø0.45: 120066 (Approx. 0.01 N)* 4 *1 Only VL-50S is available. *2 Refer to page G-18 for details of the RS link. *3 Not applicable to VL-50-100-B and VL-50S-100-B *4 V alues in parentheses indicate the measuring force of a 0.01 N model fitted with the respective optional points

Reflection type linear encoder

Stroke

51.5 mm (2 in) (when using a standard contact point) (0.5 + L/100) µm L=arbitrary measuring length (mm)

Measuring accuracy (20 °C)* 1

Accuracy guaranteed temperature* 2 Repeatability* 1 Measuring force* 1

20±1 °C

=0.05 µm

0.15 N* 3

1 N* 3

0.15 N* 3

1 N* 3

0.01 N

0.01 N

Measurement

Approx. 2 mm/s (0.08 in/s) or 4 mm/s (0.16 in/s) (changeable by parameter)

Feed speed

Fast feed

Approx. 8 mm/s (0.3 in/s)

ø3 mm carbide tipped (fixing screw: (M2.5x0.45), standard contact point: 901312

Contact point Measuring table

ø100 (ceramic, grooved, removable)

Input

Foot switch input (when optional foot switch is used) External Control Digimatic output/RS-232C output (changeable by parameter) AC adapter: 357651 , Grounding wire: 09CAA985 , AC cable: 02ZAA010 Hex wrench (2 pcs. for fixing contact point and for removing fixing bracket) Max. 18W (12V, 1.5A)

Output

Rating Power consumption

Standard Accessories

*1 Normal measurement using standard contact point. *2 Under less temperature change, and hot or cold direct air flow should be avoided. *3 0.15 N, 1 N types are factory-installed option. Note: Motor life is approximately 100,000 operations, after which replacement is advisable. This maintenance factor is particularly important to bear in mind when the machine is used frequently, such as on a production line.

DIMENSIONS 318-221-10A

318-226-10A

Unit: mm

85

150

103.2

Rear of display unit

Scale unit

Connecting cable 2 m

ø32

Stylus

Display unit

80.2

Clamp screw for measuring table Stylus

56

(26.2)

14.8

80

Measuring table ø100

Display unit Operation switch

4×M4 depth 10 (Dedicated stand-mounted dimensions)

Refer to the Litematic Brochure ( E13006-US ) for more details.

50.2

186.4

202

300

G-20

Powered by