Data Collection Software for Excel USB-ITPAK V3.0
Example of measurement using the U-WAVE Series wireless communication system
<Data sorting of individual measurements>
Data from multiple Digimatic gages ( U-WAVE-T , U-WAVE-TM/TC ) sent to separate Excel sheets Measurement data from multiple measuring instruments can easily be sorted and inserted into the respective sheets in Excel.
A
Entry point can be specified per U-WAVE ID . • Specifying an Excel file • Specifying data input cells (example: A1: C3) • Specifying cursor move
USB-ITPAK V3.0 / V2.1 (Individual measurement)
ID=99
ID=98
A C 1 2.341 2.274 2.007 B
B C 1 2.341 2.274 2.007 2 2.039 196.3 2.274 3 1.996 Sheet 98 A B
A C 1 2.341 2.274 2.007 B 2 2.039 1.963 3 1.996 2.152 Sheet 99
A C 1 2.341 2.274 2.007 2 2.039 196.3 2.274 3 1.996 Sheet 00
2 2.039 1.963 3 1.996 2.152 Sheet 01
The last data input may be canceled by pressing and holding the data switch
ID=01
ID=00
U-WAVE-T
U-WAVE-T
U-WAVE-T
U-WAVE-T
U-WAVE-R
………
Up to 100 gages can be handled by one U-WAVE-R unit Note: U-WAVE-TM / TC is also applicable.
ID=00
ID=01
ID=98
ID=99
<timer input + measurement date/time display during simultaneous measurement>
Automatically obtains displacement data in a certain input interval
The event-driven operation function of U-WAVE enables data collection at a preset time by timer.
USB-ITPAK V3.0 / V2.1 simultaneous measurement + timer input (example: 5 s interval) A B C D E F G 1 Displacement 1) Displacement 2) Displacement 3) Displacement 4) Displacement 5) Displacement 6) Measurement date/time 2 0.281 0.162 0.121 0.051 0.011 −0.001 2013/4/1 7 30 00 3 0.279 0.152 0.133 0.064 0.018 −0.003 2013/4/1 7 30 05 4 0.265 0.149 0.142 0.089 0.021 −0.007 2013/4/1 7 30 10
5 s 5 s 5 s
5 6
The input interval can be arbitrarily set by 0.1 second intervals up to 24 hours. If a smaller value than the data loading time is set, the actual measurement time will be the input interval. With U-WAVE , an error (no data) may occur if less than 0.5 seconds is set for the input interval. This is because the data request signal is issued before the data comes in, based on the event driven data refresh interval that is set to 0.5 seconds (fixed).